A GTEM (Gigahertz Transverse Electro Magnetic) cell is a test site for efficiently performing both radiated immunity and emissions testing in a single, controllable and shielded environment. Compared to other test sites, GTEM testing is faster with high accuracy and excellent reproducibility. In principle, the GTEM cell is a coaxial line expanding pyramidally and having an impedance of 50 Ω. At its end, the line is terminated by a combination of termination resistors and RF absorbers designed and constructed to match the above mentioned impedance. The GTEM 2000 has a maximum septum height of 2000 mm and is suitable for emissions and immunity testing.
Parameter | Value |
---|---|
Max. septum height |
2000 mm |
Septum height at marker position |
1732 mm |
Dimension (LxWxH) |
8.95 x 4.62 x 3.24 m |
Weight |
approx. 2500 kg |
Height H1 of cell corpus |
3.05 m |