A GTEM (Gigahertz Transverse Electro Magnetic) cell is a test site for efficiently performing both radiated immunity and emissions testing in a single, controllable and shielded environment. Compared to other test sites, GTEM testing is faster with high accuracy and excellent reproducibility. In principle, the GTEM cell is a coaxial line expanding pyramidally and having an impedance of 50 Ω. At its end, the line is terminated by a combination of termination resistors and RF absorbers designed and constructed to match the above mentioned impedance. The GTEM 1750 has a maximum septum height of 1750 mm and is suitable for emissions and immunity testing.
Parameter | Value |
---|---|
Max. septum height |
1750 mm |
Septum height at marker position |
1516 mm |
Dimension (LxWxH) |
7.95 x 4.10 x 2.90 m |
Weight |
approx. 1800 kg |
Height H1 of cell corpus |
2.75 m |
AMETEK CTS GmbH
Sternenhofstr. 15
4153 Reinach, Switzerland
Tel.+ 41 61 204 41 00
marketing.cts@ametek.com
Managing Directors:
Brian Huntsman
Astrid Wälchli Schreiber
VAT-ID: CHE-179.452.414
Register of companies: CH-280.4.015.069-3